Abstract

A new endstation has recently been commissioned at 6-m toroidal grating monochromator beamline to perform high-resolution photoemission studies in the photon energy range of 10–180 eV. The new endstation is equipped with Phoibos-150 analyzer with a channelplate and charge coupled device (CCD) camera readout to measure the kinetic energy of the electron and its emission angle simultaneously. The initial energy resolution test of the analyzer from He-I discharge lamp excited Argon 3p gas phase spectra reveals ∼20 meV full-width-half-maximum (FWHM). With this new set-up, we have investigated the electronic structure of FeAl(1 1 0). Preferential sputtering of clean FeAl(1 1 0) surface results in depletion of Al in the near surface region. The surface forms reconstructed phases at different annealing temperature regions. The electronic structure of these phases is presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.