Abstract

At the University of Hong Kong’s positron beam two related positron defect spectroscopies are being developed, namely, positron deep level transient spectroscopy (PDLTS), and coincidence Doppler broadening spectroscopy. Both of these spectroscopies require the use of high purity Ge detectors and the standard high gain nuclear instrumentation amplifiers. In this paper we review some of the difficulties that we have encountered in making such measurements satisfactorily in one of the busiest cities of the world. Attempts, such as using battery power supply, opto-isolating from local mains, connecting to an independent ground and adding a spectrum stabilizer, have been made to overcome the noise problem. However, the results are not particularly promising. The reasons why are discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.