Abstract

In this study, we determine the electron tomography (ET) resolution for microns-thick specimens by experiment in the ultra-high voltage electron microscope. A tilt series of projection images of a tilted 8μm thick epoxy-resin film are first acquired. Tomographic reconstructions are then calculated and the resolution is evaluated with the Fourier shell correlation method. The ET resolution of 32nm is achieved under the condition of 2MV accelerating voltage. We also demonstrate that some high tilt angle projections may be little useful for improving the final ET resolution because of the corresponding poor image qualities. These results are helpful to understand the possibility and limitation of ET applications in microns-thick specimens.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.