Abstract

In transmission electron microscopy the natural contrast of polymers is very low. Thus the contrast has to be enhanced by staining with heavy metals. The resolution is limited by the size of the staining particles and by the fact that electrons with different energy are focused in different image planes due to the chromatic aberration of the magnetic lenses. The integration of an electron energy loss spectrometer into the optical coloumn of a transmission electron microscope offers the possibility to use monoenergetic electrons and to select electrons with a certain energy for imaging. Thus contrast and resolution are enhanced. By imaging only electrons with an element specific energy loss the element distribution in the sample can be obtained. In addition, elastic bright field images and diffraction patterns yield excellent resolution. Some applications of the method on multicomponent polymer materials are discussed.Bulk polymer samples were prepared by ultramicrotoming at room temperature or well below the glass transition temperature. Very thin films for the direct observation of the structure in semicrystalline polymers were obtained by melt-spinning. Specimens were examined with a ZEISS CEM 902 operated at 80 kV.