Abstract

Design, characterization, and selected applications of a novel electron detector dedicated to conventional perpendicular- and low-angle-incidence conversion electron Mössbauer spectroscopy are presented. The setup is suitable for varying the incident angle and external magnetic fields on Mössbauer source and absorber. Test experiments were performed on alpha-(57)Fe films using a conventional single-line (57)Co(Rh) and magnetically split, (57)Co(alpha-Fe) Mössbauer sources. The integral "blackness effect" in conversion-electron Mössbauer spectra of (57)Fe isotope-enriched absorbers is demonstrated and shown to be pronounced at shallow angles of incidence. In order to determine the alignment and sign of the hyperfine field in an isotope-enriched absorber, the blackness effect is accounted for in a semiempirical way by using single-line source/absorber experimental relative intensities determined independently. This method works with high accuracy for linear polarimetry; however it is only a rough approximation in the case of nearly circular polarimetry.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.