Abstract

A fluorescence correlation spectroscopy (FCS) setup is built with an electron multiplying charge-coupled device camera. Although the instrument has a limited time resolution of 4 ms, compared to 0.1-0.2 mus for common instruments using avalanche photodiodes, it allows multiplexing of FCS measurements, has a software-adjustable pinhole after data collection, performs flow speed as well as flow direction measurements in microchannels and could be used to do spectral FCS. Measurements are performed on fluorescent dyes and polystyrene beads in high-viscosity media and on epidermal growth factor receptors in Chinese hamster ovary cells. Using real measurements on single spots, multiplexing of focal spots and detection elements are simulated and the results are discussed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.