Abstract

In order to introduce Electroureterography-through-cystoscope into the clinical urology, three types of electrodes were produced: ‘simple’, ‘bridge type’ and ‘U-loop’ collar electrodes. Experiments were performed with the electrodes on both animal and human subjects. 1) Satisfactory electro-ureterogram (E. U. G.) was recorded with all three types of electrodes in canine ureter. 2) E. U. G. recording with ‘simple’ and ‘bridge type’ failed in human subjects. 3) With ‘U-loop’ electrode E. U. G. was recorded satisfactorily in both human and canine ureters. The wave pattern was similar to those by Sleator and Butcher, and Shiratori et al. The action potentials measured were 0.4-1.2 mV in amplitude with an average of 0.73 mV; 0.53-1.0 sec. in duration, average 0.70 sec.; 6.8-57.5 sec. in discharge interval, average 19.3 sec.; 22.2-33.3mm/sec. in conduction velocity, average 31.1mm/sec.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.