Abstract

A method was developed and used to determine the electromechanical properties of high frequency (>20 MHz) piezoelectric strip vibrators. A nonlinear regression technique was employed to fit the impedance magnitude and phase as predicted by Mason’s model to measured values. Results from experimental measurements on 30 MHz array elements supported by an attenuative backing indicated degraded performance when compared to values predicted from the electromechanical properties measured at low frequency. This degradation may be attributed to damage incurred during fabrication and grain size effects, with a fine grain sized material providing superior relative performance. This technique may be used in the evaluation and comparison of different fabrication processes and materials for high frequency medical imaging arrays.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.