Abstract

Electromagnetic interference (EMI) to critical care medical devices has been reported by various groups. Previuos study demonstrated that infusion pumps are susceptible of false alarm buzzing and block of infusion, when exposed to various EMI sources. Aim of this paper is to investigate the changes in the risk of EMI from the estimates of our previous 2005 survey and to extend the EMI risk assessment to newer telecommunication products: DECT phones and WiFi terminals. With regards to GSM phones, compare to the results obtained in 2005, we observed a decrease in the rate of failure (from 58% to 30%). From our findings, the use of WiFi and DECT does not pose a real risk to infusion systems.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.