Abstract

There have been a number of reports of electromagnetic interference (EMI) on electronic medical equipment caused by mobile telecommunication systems. In Japan, the use of the personal handy-phone system (PHS) has greatly expanded within urban areas, PHS handsets transmit EM signals at a frequency of 1.9 GHz and have a peak radiated power of 80 mW. This power level is lower than that of other mobile telecommunication systems. Two studies were carried out. One was to determine whether or not PHS interferes with electronic medical equipment in hospitals. We observed no EMI on electronic medical equipment when the PHS handset was in either the speaking mode or on standby. The second study was to observe EMI from 1.9-GHz signals at several radiation power levels. Although EMI was not observed at the radiated peak power of the PHS handset, EMI on some of the tested equipment was observed when the radiated power was ten or more times higher than that of the PHS handset.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.