Abstract

The increasing ageing population is leading to a wide-scale demand for health-state monitoring by a wireless body area network (BAN). Wireless BAN needs each vital sensor to act as a wearable device for collecting blood pressure, electrocardiogram (ECG), electroencephalogram and so on in daily life. On the other hand, wireless power transfer is also getting into our daily life because of its convenience, which suggests a potential electromagnetic interference (EMI) problem on the wearable devices in healthcare and medical BAN. In this study, the authors quantitatively evaluated the EMI on wearable ECG for 6.8 MHz wireless power transfer system. They employed electromagnetic field analysis technique to derive the common-mode voltage between the human body with a wearable ECG and the ground plane, and circuit simulation or measurement to derive the interference voltage at the wearable ECG output. The result first time gave a quantitative evaluation for EMI of wireless power transfer on wearable ECG. The approach is also available to be applied to EMI evaluation of other wearable devices in healthcare or medical applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.