Abstract

There has been concern regarding the effect of electromagnetic interference (EMI) on medical equipment from portable wireless communication devices in hospitals in Hong Kong. The latest third-generation (3G) mobile communication system was fully implemented in 2004. As a result, an investigation to examine the EMI issues surrounding medical equipment, due to interference from communication devices, including GSM900, PCS1800, and 3G digital mobile phones, has been carried out. A statistical analysis has shown that the cumulative distribution function of base station signals within the hospital's environment follows the Weibull distribution for all three mobile phone systems in Hong Kong and that the percentage of failed medical equipment increases with a correlation of an increase in the radiated power level of the mobile phone. This is a direct result of a weakening in base station signals. The EMI effect from 3G mobile phones is also found to be four to eight times lower than that of the GSM900 and PCS1800, in locations inside the hospitals with weak base station signal strength, such as a basement or in operation theaters.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.