Abstract
In a previous paper, a reciprocity calibration formula different from the Levy–Bouche’s is derived [L.-F. Ge, Proceedings of the XIII IMEKO World Congress, 2606–2610 (1994)]; the latter has been regarded as a standard method [ISO 5347-0, 1987; ISO/DIS 5347-20, 1990]. The new formula is S=(RT/jω)1/2 (V/ms−2), where T is the mean of the transfer admittance ratio, Tn=Mn/(Yn−Yo) and R is the voltage ratio for the typical three-transducer-two-experiment procedure, in which an electrodynamic transducer serves as reciprocal one. This method is suited to the low- and mid-frequency ranges, and most accurate at 100–1000 Hz. This paper reveals that a coefficient of variation, i.e., ratio of standard deviation of Tn to its mean value, can be viewed as a reciprocity index of such an electrodynamic calibration system including two transducers and a solid mass medium. An index smaller than 0.5% indicates that the system is better reciprocal, and can be calibrated more accurately. At frequencies higher than about 3000 Hz, nonlinear problems, such as stress wave effect and resonance, become pronounced, and should be considered; a better way would be to apply other more available methods, such as the laser interferometry and piezoelectric reciprocity method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.