Abstract

Simultaneous optical and voltammetric detection of bioactivated genotoxicity is reported for the first time employing ultrathin films of DNA, model metabolic enzymes, and electrochemiluminescence (ECL) generating metallopolymer [Ru(bpy) 2PVP 10] 2+ on pyrolytic graphite (PG) electrodes. Cytochrome P450 cam and myoglobin were used as model monoxygenase enzymes to mimic in vivo processes. Sensor film growth and component amounts were monitored using a quartz crystal microbalance (QCM). Subsequent to the enzyme reaction, DNA damage in the sensor films was measured simultaneously using a simple apparatus combining a standard voltammetry cell coupled with an optical fiber and photomultiplier tube. The model enzyme reaction converted styrene to styrene oxide, which reacts with DNA nucleobases. ECL and SWV signals increased with enzyme reaction time on the scale of several min, and provided relative enzyme turnover rates for DNA damage suitable for toxicity screening applications. Within 1 min, the sensor detects ∼3 damaged bases per 10,000 DNA bases using this simultaneous detection.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.