Abstract

An electrochemiluminescence (ECL) biosensor was fabricated for the evaluation of prostate specific antigen (PSA). The sensor was developed by successively modifying glassy carbon electrode (GCE) electrodes with CdS/Chito/g-C3N4 nanocomposites and DNA1 was labeled at the 5' end with thiol. The aptamer DNA was labeled at the 3' end with a quencher ferrocene (Fc) was ligated to DNA1 by the principle of complementary base pairing. In the absence of PSA, the ECL intensity signal is effectively quenches through the energy transfer and photoexcitation electron transfer between CdS/Chito/g-C3N4 emitter and quencher Fc. After incubation with target PSA, the aptamer DNA interacts with PSA and then moved away from the electrode surface together, which will recover the ECL intensity. Under the optimal conditions, the ECL intensity increases linearly with the logarithm of PSA concentration in the range of 1pg·mL-1 to 100ng·mL-1, and the detection limit is 0.14pg·mL-1 (S/N = 3). The biosensor has been successfully applied to the determination of PSA in serum sample. Graphical abstractSchematic representation of the electrochemiluminescence sensor based on a CdS/chitosan/g-C3N4 nanocomposite, which can be applied to the determination of prostate specific antigen in serum.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.