Abstract

AbstractHerein we report the study of electrochemiluminescence (ECL) generation by tris(2,2′‐bipyridyl)ruthenium (Ru(bpy)32+) and five tertiary amine coreactants. The ECL distance and lifetime of coreactant radical cations were measured by ECL self‐interference spectroscopy. And the reactivity of coreactants was quantitatively evaluated in terms of integrated ECL intensity. By statistical analysis of ECL images of single Ru(bpy)32+‐labeled microbeads, we propose that ECL distance and reactivity of coreactant codetermine the emission intensity and thus the sensitivity of immunoassay. 2,2‐bis(hydroxymethyl)‐2,2′,2′′‐nitrilotriethanol (BIS‐TRIS) can well balance ECL distance‐reactivity trade‐off and enhance the sensitivity by 236 % compared with tri‐n‐propylamine (TPrA) in the bead‐based immunoassay of carcinoembryonic antigen. The study brings an insightful understanding of ECL generation in bead‐based immunoassay and a way of maximizing the analytical sensitivity from the aspect of coreactant.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.