Abstract

The paper presents the validation of an electrochemical procedure for on-site Hg2+ ions determination in wastewater samples using a modified carbon screen-printed electrode (SPE) with a complexing polymeric film based on poly(2,2'-(ethane-1,2-diylbis((2-(azulen-2-ylamino)-2-oxoethyl)azanediyl))diacetic acid) (polyL). Using metal ions accumulation in an open circuit followed by anodic stripping voltammetry, the SPE-polyL electrode presents a linear range in the range of 20 µg/L to 150 µg/L, with a limit of detection (LOD) = 6 µg/L, limit of quantification (LOQ) = 20 µg/L, and an average measurement uncertainty of 26% of mercury ions. The results obtained in situ and in the laboratory using the SPE-polyL modified electrode were compared with those obtained by the atomic absorption spectrometry coupled with the cold vapor generation standardized method, with the average values indicating excellent recovery yields.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.