Abstract
Multiplexing assay of biomarkers at the point-of-care is an elusive goal for molecular diagnostics. Here, we report an electrochemical (EC) sensor for oral cancer detection based on the simultaneous detection of two salivary biomarkers: interleukin (IL)-8 mRNA and IL-8 protein. Under the multiplexing mode, the limit of detection of salivary IL-8 mRNA reaches to 3.9 fM and 7.4 pg/mL for IL-8 protein in saliva. Multiplex assay of these 2 biomarkers directly from 28 cancer and 28 matched control saliva samples shows significant difference between the two groups. From the receiver operating characteristic analysis, the EC sensor yields around 90% sensitivity and specificity for both IL-8 mRNA and IL-8 protein, which are very close to the data measured by traditional assays (ELISA and PCR) with the same group of saliva. Combined IL-8 mRNA and protein show better AUC compared with single biomarker. We show, for the first time, concurrently multiplexing detection of salivary mRNA and protein biomarkers using point-of-care EC sensor.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.