Abstract

A novel electrochemical sensor for cholesterol (CHO) detection based on molecularly imprinted polymer (MIP) membranes on a glassy carbon electrode (GCE) modified with multi-walled carbon nanotubes (MWNTs) and Au nanoparticles (AuNPs) was constructed. p-Aminothiophenol (P-ATP) and CHO were assembled on the surface of the modified GCE by the formation of Au–S bonds and hydrogen-bonding interactions, and polymer membranes were formed by electropolymerization in a polymer solution containing p-ATP, HAuCl4, tetrabutylammonium perchlorate (TBAP) and the template molecule CHO. Cyclic voltammetry (CV) and differential pulse voltammetry (DPV) measurements were used to monitor the electropolymerization process and its optimization, which was further characterized by scanning electron microscopy (SEM). The linear response range of the MIP sensor was between 1×10−13 and 1×10−9molL−1, and the limit of detection (LOD) were 3.3×10−14molL−1. The proposed system has the potential for application in clinical diagnostics of cholesterol with high-speed real-time detection capability, low sample consumption, high sensitivity, low interference and good stability.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.