Abstract
Uranium may be accumulated at an anodized glassy carbon (GC) electrode at −0.15 V and stripped at +1.15 V, allowing electrochemical preconcentration and matrix elimination on-line with ICP-MS detection. Analyte levels ranging from 20 μg/L to subnanogram per liter were determined in matrix-free standards with good linearity and precision. Detection limits (3sy/x/slope) determined from standard addition plots were 0.12 ng/L with 10-min accumulation time, permitting the minor isotopes of uranium to be determined with good precision at total uranium concentrations as low as 5.0 μg/L. High levels of salt matrix moderately suppressed uranium deposition efficiency, and therefore stripping peak height, but could be compensated for by extended accumulation times. Matrix elimination of up to 20 000 mg/L Na was complete with just a 90-s cell washout time. Uranium was quantified in a certified seawater matrix, NASS-4 (certified uranium level 2.68 ± 0.12 μg/L), at several levels of dilution: recoveries ranged from 10...
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.