Abstract

AbstractThis paper describes a layer‐by‐layer (LBL) self‐assembly process of chitosan (CTS) and gold nanoparticles (Au) on the pretreated negatively charged glassy carbon (GC) electrode to fabricate electrochemistry immunosensor with a nontoxic biomimetic interface, which provided an environment similar to a native system and allowed more freedom in orientation for immobilization of carcinoembryonic antibody (anti‐CEA) to monitor carcinoembryonic antigen (CEA). UV‐vis spectroscope, atomic force microscopy (AFM), and cyclic voltammetric (CV) measurements were used to follow the multilayer film formation. The performance of the biominetic interface and factors influencing the assay system were investigated in detail. The differential pulse voltammetry (DPV) current response is used for the CEA concentration assay. The dynamic range was from 0.50 to 80.00 ng mL−1 with a detection limit of 0.27 ng mL−1 at 3σ. In addition, the experiment results indicate that immobilization described in this proposed method exhibits a good sensitivity, selectivity, and stability.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.