Abstract

Abstract A novel pyranose oxidase (PyOx) biosensor based on gold nanoparticles (AuNPs)–polyaniline(PANI)/AgCl/gelatin nanocomposite has been developed for the glucose detection. PyOx was immobilized on the surface of glassy carbon electrode (GCE) via the nanocomposite matrix. The electrode surface was imaged by scanning electron microscopy (SEM). Amperometric detection of the consumed oxygen during the enzymatic reaction was monitored at −0.7 V. After optimization studies, analytical characterization of the biosensor was carried out. The linear response of the AuNPs–AgCl/PANI/gelatin modified PyOx biosensor is found to be from 0.05 to 0.75 mM glucose with the equation of y = 2.043x + 0.253; R2 = 0.993. Finally, proposed biosensor was used to analyze glucose content in real samples. Obtained data from the biosensing system was compared with a commercial enzyme assay kit based on spectrophotometric Trinder reaction as a reference method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.