Abstract

Deoxyribonucleic acid (DNA) was electrochemically deposited on a carbon ionic liquid electrode to give a biosensor with excellent redox activity towards paraquat as shown by cyclic voltammetry and differential pulse voltammetry. Experimental conditions were optimized with respect to sensing paraquat by varying the electrochemical parameters, solution pH, and accumulation time of DNA. Under the optimized conditions, a linear relation exists between the reduction peak current and the concentration of paraquat in the range from 5 × 10−8 mol L−1 to 7 × 10−5 mol L−1, with a detection limit of 3.6 × 10−9 mol L−1. The utility of the method is illustrated by successful analysis of paraquat in spiked real water samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.