Abstract

A novel and versatile method, based on a membrane-free enzyme electrode in which both the enzyme and a mediator protein are entrapped in a gelatine hydrogel was developed for the fabrication of biosensors. As a proof of principle, we prepared a hydrogen peroxide biosensor by successfully entrapping both horse heart cytochrome c (HHC) and Saccharomyces cerevisae cytochrome c peroxidase (CCP) in a gelatin matrix which is immobilized on a gold electrode. This electrode was first pretreated with 6-mercaptohexanol. The biosensor displayed a rapid response and an expanded linear response range from 0 to 0.3 mM (R = 0.987) with a detection limit of 1 × 10 − 5 M in a HEPES buffer solution (pH 7.0). This method of encapsulation is now further investigated for industrial biosensor applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.