Hydroperoxy derivatives of phosphatidylcholine (PC) and phosphatidylethanolamine (PE) can be separated from their respective phospholipids by reverse-phase high performance liquid chromatography (HPLC). However, ultraviolet absorption due to conjugated diene cannot detect the hydroperoxy group. In this work, an electrochemical (EC) detector was first applied to the analysis of hydroperoxy phospholipids. Both the PC and PE hydroperoxides from rat liver were reduced quantitatively by the glassy carbon electrode at -300 mV vs Ag/AgCl. Since neither the hydroxy derivatives nor unoxidized phospholipids showed any response, it would seem this technique can be used to distinguish phospholipid hydroperoxides from their hydroxy derivatives. Thus, the reverse phase HPLC-EC detection method is proposed for the specific analysis of hydroperoxy phospholipids in biological tissues.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.