Abstract

Electrochemical detection of hydrogen peroxide (H2O2) at a platinum-deposited boron doped diamond was studied for an application in melamine strip test. Deposition of Pt particles at a boron-doped diamond surface was performed using cyclic voltammetry method. The method was also used for electrochemical measurements. The oxidation–reduction peak currents of H2O2 at +0.6V and −0.1V (vs. Ag/AgCl), respectively were found to be linear in the concentration range of 0.05–20mM (R2=0.99). Linearity was also observed to be dependent on horseradish peroxidase (HRP) concentration in the constant concentration of H2O2. The method was then applied in immunochromatographic strip test for melamine detection to determine the activity of HRP labeled at melamine antibody. Taking an assumption that the activity of HRP was equivalent to melamine concentration, linearity of the current responses to the concentrations of standard melamine solutions in the range of 5–100μg/ml could be achieved. Limit of detection of 0.51μg/ml was estimated. The obtained results demonstrated that the designed biosensor for detection of H2O2 could be applied for the quantitative and selective detection of melamine.