Abstract

In this work, a sandwich-type electrochemical immunosensor for carcinoembryonic antigen (CEA) detection has been constructed and tested. Unlike many other sensors using external electrochemical species in the electrolyte to generate an electrochemical signal, a ferrocene derivative has been integrated into the design of the sensor to provide an internal reporting system, allowing detection of CEA in buffers and biological samples. Gold nanoparticles, which have been used to increase the conductivity of sensing surfaces, also carry immobilized secondary anti-CEA and a ferrocene derivative. The shelf life testing of the sensor shows good performance after storage for 4 weeks. The sensor has been calibrated against different concentration of the target protein using square wave voltammetry. The calibration curve has been obtained in the range of 0.05–20ngmL−1, and the detection limit for CEA is ~ 0.01ngmL−1. The capability of the immunosensor has been verified by performing detection of CEA in human serum samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.