Abstract

A reduced graphene oxide (rGO)/Fe3O4 composites modified screen-printed carbon electrode (SPCE) was used to determinate As(III) in a HAc-NaAc buffer solution. The rGO/Fe3O4 composites were prepared by a simple and one-pot synthesis method, and characterized by X-ray diffraction (XRD), field emission scanning electron microscopy (SEM), FT-IR and Raman spectra. The electrochemical behaviors of the composite electrode were characterized by cycle voltammetry and electrochemical impedance spectroscopy. The experimental parameters, such as supporting electrolyte, solution pH, deposition potential, deposition time were optimized, respectively. The calibration curve for the detection of As(III) in the range of 2 to 20 ppb was I = -4.495 + 1.922C, with a coefficient of 0.994. The sensitivity was 1.922 μA/ppb, and with about twice Fe3O4 modified SPCE, a detection limit as low as 0.3 ppb was achieved. The proposed electrode was validated by analyzing the As(III) content in real water samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.