Abstract
The study presents results of electrochemical corrosion tests of extruded magnesium alloy WE43 used for stents in cardiology. Corrosion tests were performed in artificial plasma with application of the system for electrochemical tests VoltaLab PGP201. Potentiodynamic tests enabled to register polarisation curves. Scanning microscope were used for assessment of the surface of tested alloy after corrosion tests. It was proved that the alloy is susceptible to pitting corrosion. Options of magnesium alloy WE43 application in cardiology are connected with the need for application of protective layers on implants made of the tested alloy.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.