Abstract
The electrochemical characterization of boron-doped polycrystalline diamond thin-film (BDF) electrodes was studied using the anodic scan after concentrating lead in 0.1 mol/L KCl - 41 micromol/L Hg(NO(3))(2) and 0.1 mol/L KNO(3) - 0.01 mol/L HNO(3) - 41 micromol/L Hg(NO(3))(2); accumulation voltage was -0.90 V. The results obtained were compared with those given by glassy carbon (GC) electrodes and proved that the BDF electrodes offered high sensitivity, good precision and extreme stability over a 2-month period. These electrodes provided good resolving power for the determination of lead and cadmium and gave satisfactory results in the analysis of a pure water sample.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.