Abstract

Guanosine-5′-monophosphate (GMP) was investigated the electrochemical behaviors based on solid-phase extractionon (SPE) at Cu-Mg-Al hydrotalcite-like compound (HTLC) modified glass carbon electrode. Cu-Mg-Al hydrotalcite-like compound (HTLC) was proved as a new sorbent for SPE of GMP, which showed an irreversible adsorption oxidation process on the HTLC/GCE with the oxidation peak potential located at 1.15 V (vs. SCE) in a pH 5.0 acetate buffer solution. Influencing factors of the electrochemical behavior of GMP on the HLTC/GCE were optimized and kinetic parameters were calculated. Under the optimal conditions, with differential pulse voltammetry (DPV), a linear relationship was obtained between the oxidation peak current and the GMP concentration in the range from 1.0 × 10− 6 to 8.0 × 10−4 mol L−1 with the detection limit as 5.0 × 10−7 mol L−1 (signal-to-noise ratio of 3). The modified electrode surface has very good reproducibility and stability.

Full Text
Published version (Free)

Talk to us