Abstract

We present, for the very first time, the fabrication and electrochemical characterization of a paper-based experimental platform for dengue virus analysis. The paper-based device incorporates a screen-printing technology with the help of black carbon conductive ink. The paper-based device utilizes two styles of electrode setups, i.e., the two-electrode system and three-electrode system, and both setups effectively detected the dengue virus with an LOD of 0.1 µg/mL; however, these paper electrodes exhibit various current ranges, and the created sensor was encompassed and compared in this research based on current response. It is observed that the three-electrode system has a substantially higher current range, ranging from 55.53 µA to 322.21 µA, as compared to the two-electrode system, which has a current range of 0.85 µA to 4.54 µA. According to this study, the three-electrode system displayed a good range of current amplification that is roughly 50 times higher than the two-electrode system, which had a weak current response. As a result, the three-electrode method has emerged as a viable option for the very sensitive detection of the dengue virus, as well as for the diagnosis of other diseases.