In this paper, corrosion damage in the modular femoral hip prosthesis which is retrieved from the physiological environment for 58 months after insertion, has been analyzed with the parameter of the square root of projected area of corrosion pit from the view point of fracture mechanics. The effects of the pitting damage on fatigue properties of Ti-6Al-4V alloy was discussed based on the previous experimental result. Accelerated laboratory corrosion testing method which simulates the physiological corrosion environment has been developed on the basis of the results of corrosion test in simulated environment with Ti-6Al-4V alloy. The empirical formula which represents the relationship of applied current, corrosion time and maximum size of corrosion pit (√(area)max) has been proposed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.