Abstract

We report a case of electrocautery-induced pacemaker failure that resulted in asystole in a 15-year-old girl scheduled for cardiac surgery. Her pacemaker was converted to the asynchronous mode the night before surgery. Electromagnetic interference from the unipolar electrocautery caused a reduction in the battery voltage, which allowed the digital circuitry, but not the voltage control oscillator (VCO), to work properly. Eventually the battery current drain caused VCO "lock-out," and pacemaker and battery failure. This report demonstrates that electrocautery-induced pacemaker failure can occur, even after conversion to asynchronous mode.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.