Abstract

A novel modified glassy carbon electrode containing Ni(Me2(CH3CO)2[14]tetraenoN4) complex was used as an electrocatalytic sensor for the determination of L-ascorbic acid in pH = 6.6. The peak potential shifted to negative by 205 mV compared with that for a bare electrode in cyclic voltammograms. The calibration curve was linear up to 6.2 x 10(-3) M with a detection limit 3.1 x 10(-7) M and an RSD% better than 2.47%. This newly modified electrode was applied to commercial pharmaceutical tablets, injections and foods. The obtained results were identical to those obtained by the classical 2,6-dichlorophenolindophenol method.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.