Abstract

AbstractIn this research, we report the preparation of a high sensitive voltammetric sensor for electrocatalytic determination of hydroxylamine (HX) in the presence of thiosulfate (TS). We describe the synthesis and characterization of CdO nanoparticles (CdO/NPs) with transmission electron microscopy (TEM) and X‐ray diffraction (XRD) methods. At an optimum condition, the two peaks are separated ca. 0.58 and 0.92 V for HX and TS at the surface of the proposed sensor. Square wave voltammetry (SWV) of HX exhibited two linear dynamic ranges with a detection limit (3σ) of 0.06 µmol L−1. The propose sensor was used for determination of HX in water samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.