Abstract

AbstractAn electrochemical sensor for the simultaneous determination of Cd(II) and Pb(II) by square wave anodic stripping voltammetry (SWASV) in bivalve mollusks using a glassy carbon electrode modified with electrochemically reduced graphene oxide has been developed. The modified surface was characterized by cyclic voltammetry, high resolution scanning electron microscopy (HR‐SEM), and Raman spectroscopy. The optimum conditions were optimized and a linear range was observed from 15–105 μg L−1 with a limits of detection of 15 μg L−1 for Cd(II) and Pb(II). The methodology was validated and applied in different samples of commercial bivalve mollusks with satisfactory results. The high conductivity and greater surface area of the modifying agent improves the preconcentration capacity of the electrochemical sensor, allowing to develop a simple, rapid and sensitive analysis in the detection of lead and cadmium in marine resources.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.