Abstract

This work reports the construction of a novel nanostructured immunosensor for detection of the troponin I biomarker (cTnI). Anti-troponin I antibody was anchored on the modified graphite electrode with reduced graphene oxide and polytyramine for detection of troponin I in serum samples. The performance of the electro-immunosensor was evaluated by differential pulse voltammetry. The immunosensor presented a wide work range, from 4ng mL-1 to 4pg mL-1 , whose detection limit (4pg mL-1 ) is significantly lower than the basal level in human serum, and maintained 100% of response after 30 days of storage. Moreover, the immunosensor showed good selectivity for detection of cTnI in real sample containing interfering substances and specificity of response to cTnI in the serum of healthy and sick patients, and demonstrated the possibility of reuse for two consecutive analyses, in addition to using a simplified and inexpensive platform when compared to other devices, demonstrating them excellent potential for application in diagnosis in the early stages of acute myocardial infarction.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.