Abstract

A new approach to radiant power measurements is described. A continuously tunable cw dye laser was used to measure the absolute spectral response of a silicon photodiode and narrow band−pass filter by comparison with an electrically calibrated pyroelectric detector. The filtered photodetector was then used to measure the spectral power density from a standard lamp that had been calibrated by the classical technique. The agreement of better than 1% between the two measurements is well within the uncertainties identified with each measurement individually, of the order of 1%. A number of advantages in the new technique are identified.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.