This paper developed an electrical micro flow cytometry to realize leukocyte differentials leveraging a constrictional microchannel and a deep neural network. Firstly, purified granulocytes, lymphocytes or monocytes traveled through the constrictional microchannel with a cross-sectional area marginally larger than individual cells and produced large impedance variations by blocking focused electric field lines. By optimizing key elements (e.g., normalization, learning rate, batch size and neuron number) of the recurrent neural network (RNN), electrical results of purified leukocytes were analyzed to establish a leukocyte differential system with a classification accuracy of 95.2%. Then the leukocyte mixtures were forced to travel through the same constrictional microchannel, producing mixed impedance profiles which were classified into granulocytes, lymphocytes and monocytes based on the aforementioned differential system. As to the classification results, two leukocyte mixtures from the same donor were processed, producing comparable classification results, which were 57% versus 59% of granulocytes, 37% versus 34% of lymphocytes and 6% versus 7% of monocytes. These results validated the established classification system based on the constrictional microchannel and the recurrent neural network, providing a new perspective of differentiating white blood cells by electrical flow cytometry.