Abstract

The Flexible Imaging Diffraction Diagnostic for Laser Experiments (FIDDLE) is a newly developed diagnostic for imaging time resolved diffraction in experiments at the National Ignition Facility (NIF). It builds on the successes of its predecessor, the Gated Diffraction Development Diagnostic (G3D). The FIDDLE was designed to support eight Daedalus version 2 sensors (six more hCMOS sensors than any other hCMOS-based diagnostic in NIF to date) and an integrated streak camera. We will review the electrical requirements, design, and performance of the electrical subsystems that were created to support this large number of cameras in the FIDDLE. The analysis of the data that the FIDDLE is intended to collect relies heavily on the accurate and well-understood timing of each sensor. We report camera-to-camera timing jitter of less than 100ps rms and sensor integration times of 2.2ns FWHM in 2-2 timing mode. Additionally, diffraction experiments on the NIF produce electric fields (EMI) on the order of 1 kV/m, which have been observed to negatively impact the performance of some electrical components of the FIDDLE. We report on the results of testing hCMOS camera electronics in a similar EMI environment generated in an offline lab. We also summarize the use of a novel approach to using a vector network analyzer as an EMI leak detector to understand and reduce the negative impacts of EMI on the FIDDLE.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call