Abstract

The known effect of high current density under the perimeter of defibrillation electrodes, leading to skin damage and even severe burns in some cases, has been considered by many investigators. Two main approaches for improvement were proposed: (i) interfacing with layers of varying and high resistivity and (ii) lengthening and shaping the perimeter line. Using finite element and physical modelling, it is shown that the second approach does not yield significant improvement in the distribution uniformity. Moreover, the application of high resistivity layers is unacceptable in defibrillation. The use of a low resistance layer with a diameter covering and extending over the metal plate by at least 2.5 mm results in better uniformity. A similar effect can be obtained by recessing the metal plate in an isolating support--an approach adopted from implantable neurostimulation electrodes. These two versions can be applied in combination.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.