Abstract

Thiel-embalmed human cadaveric specimens are used widely for biomedical scientific investigation. This Letter reports electrical impedance measurements of Thiel-embalming solutions from 1 Hz to 32 MHz. Compared with other solutions studied, Thiel solution has the lowest impedance throughout the test spectrum. The electrical conductivity of a Thiel-embalmed liver sample exhibited a relatively flat frequency response from 100 to 500 kHz, which is the frequency range used for electro-surgery and radiofrequency tumour ablation (RFA). It measured 5 Sm−1 compared with 0.22 Sm−1 obtained from ex-vivo fresh pig liver. Using finite element modelling and experimental evaluation, the ablation zone obtained from the Thiel-embalmed liver sample was extremely small due to its much higher conductivity. Hence, it is concluded that Thiel-embalmed tissue cannot be used as a reliable model for RFA evaluation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.