Abstract

Rapid tests to assess the susceptibility of bacteria to antibiotics are required to inform antibiotic stewardship. We have developed a novel test, which measures changes in the impedance of a 100 nanoliter volume of bacterial suspension to determine an "electrical" minimum inhibitory concentration (eMIC). Two representative strains of Klebsiella pneumoniae, Acinetobacter baumannii, Escherichia coli, Pseudomonas aeruginosa, and Staphylococcus aureus were tested against a panel of frontline antibiotics with different modes of action (ciprofloxacin, doxycycline, colistin and imipenem, gentamicin, and ceftazidime). The eMIC measured at 1 h correlated strongly with a standard 24 h microbroth dilution MIC for all combinations of antibiotics and bacteria, allowing strains to be correctly assigned as sensitive or resistant measured in a fraction of the time.