Abstract

Electrical Capacitance Tomography (ECT) system is a non-intrusive method to detect variation of permittivity distribution in a closed pipe. ECT offers fast response, cheap and non-radiation system. However, it suffers from soft-field effect. One of the proposed solutions to overcome this problem is by applying segmentation excitation, which is to have more than one electrode excited at one time. This paper focuses on various segmentation excitations of 4-electrode combinations using COMSOL Multiphysics software. The electrical potential is recorded at the center of the pipe to study the strength of the electrical potential. The result indicates that 4-electrode adjacent configuration recorded 1.005 x 10-9V at the center of the pipe as compared to 0.838 x 10-9V for opposite configuration. This effect shows 16.6% difference in electrical potential at the center of the pipe.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.