Abstract
Percutaneous procedures to divert blood flow from one blood vessel to another can be performed with intravascular catheters but demand a method to align a crossing needle from one vessel to another. Fluoroscopic imaging alone is not adequate, and it is preferable to have a sensor on one catheter that detects the correct alignment of an incoming needle. This can be implemented by generating dipole electric fields from the crossing catheter which are detected by a receiving catheter in the target vessel and, thus, can calculate and display the degree of alignment, permitting the operator to rotate the crossing catheter to guarantee alignment when deploying a crossing needle. Catheters were built using this concept and evaluated in vitro. The results show that accurate alignment is achieved, and a successful crossing can be made. The concept is being further developed for further clinical evaluation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.