Abstract

IntroductionDetermining the normal values for acoustic radiation force impulse (ARFI) shear wave elastography of the thoracolumbar fascia (TLF) and define possible factors of influence. MethodsWe measured the shear wave velocity (SWV) in m/s and the diameter (anterior-posterior) in mm of the TLF bilateral in 267 healthy participants with the Acuson S3000™ (Siemens) using the virtual touch image quantification mode (VTIQ). The parameters were tested for correlations with the anthropometric data of the participants, between different age groups and the genders, as well as information obtained from the history, such as smoking and sporting activities. ResultsWe determined a mean SWV of 3.28 ± 0.55 m/s for the left thoracolumbar fascia and 3.44 ± 0.55 m/s for the right. The diameter on the right was 2.7 ± 0.8 mm. On the left, it was 2.7 ± 0.9 mm. Neither body mass index (BMI) nor gender had a significant effect on either of the measured parameters (p > 0.05). The same goes for regular medication, sporting activity or the consumption of alcohol (p > 0.05). The results concerning the effect of smoking and age were inconclusive as they only had a significant influence to either the right or the left side of the TLF but not on the other side. ConclusionsWe collected the normal value for ARFI shear wave elastography of the TLF in 267 healthy participants. Furthermore, neither gender, BMI, sports activity nor the consumption of alcohol affected the elasticity or the diameter of the thoracolumbar fascia.