Abstract

The aim of this study was to evaluate the effect of central corneal thickness (CCT) of keratoconic corneas on intraocular pressure (IOP) measurements as measured by four different techniques. Forty-one eyes of forty-one keratoconus patients (group 1) and fifty eyes of fifty healthy subjects (group 2) were enrolled. IOP was measured with iCare, IOPen, Goldmann applanation tonometry (GAT) and dynamic contour tonometry (DCT). CCT was measured by ultrasonic pachymetry. These data were used for statistical analysis. The mean IOPs measured by GAT, DCT, iCare and IOPen were 11.4, 14.7, 10.8, and 15.7 mmHg in group 1; and 14.2, 15.4, 15.4 and 14.3 mmHg in group 2, respectively. Between both groups, there was a statistically significant difference in iCare (p < 0.001), GAT (p < 0.001) and IOPen (p = 0.040) measurements; with no difference between DCT (p = 0.266) measurements. IOPen measurements were significantly associated with CCT (r = - 0.314; p = 0.046). IOPen seemed to be affected by CCT. IOP readings by iCare, GAT and DCT were found to be independent of CCT in keratoconic corneas.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.