This study quantified the geometric deformation of the capsular bag following implantation of various intraocular lenses (IOL) using a three-dimensional capsular bag model made of silicone caoutchouc. After implantation of 13 different IOLs (polymethylmethacrylate, silicone acrygel) into the artificial bag, the induced capsular bag deformation was measured and analyzed. The posterior space between IOL and capsule was examined by ultrasonographic biomicroscopy. Polymethylmethacrylate IOLs with C-haptic design induced a greater deformation of the capsular bag than silicone lenses; however, both types showed a larger gap posterior to the IOL than acrygel lenses. Acrygel IOLs with different haptic design revealed only minimal deformation with close contact posteriorly. The presented three-dimensional model simulates the biomechanical and geometrical parameters of the vital capsular bag. Further investigations may determine a correlation of close posterior IOL contact and lens epithelial cell progression.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.