Abstract

Eiger is a single-photon counting x-ray pixel detector being developed at the Paul Scherrer Institut (PSI) for applications at synchrotron light sources. It follows the widely utilized and successful Pilatus detector. The main features of Eiger are a pixel size of 75 × 75 μm2, high frame rate capability of 22 kHz and negligible dead time between frames of 4 μs. This article contains a detailed description of Eiger detector systems, from the 500 kpixel single-module detector to large-area multi-modules systems. The calibration and performance of the first 500 kpixel system that is in routine user operation are also presented. Furthermore, a method of calibrating the energy of single-photon counting detectors along the detector gain axis is introduced. This approach has the advantage that the detector settings can be optimized at all energies for count rate capabilities. Rate capabilities of the system are reported for energies between 6 and 16 keV.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.